Streamline High-Throughput Oligonucleotide Analysis With Ion-Pairing-Free Methods
App Note / Case Study
Published: July 10, 2025
Credit: iStock
High-throughput oligonucleotide analysis faces significant operational challenges when using traditional ion-pairing methods that require dedicated instrumentation, toxic reagents and daily mobile phase preparation.
Labs analyzing multiple oligonucleotide therapeutics often struggle with sample backlogs, instrument downtime and cross-contamination issues that slow critical research timelines.
This application note demonstrates how an ammonium bicarbonate-based approach eliminates bottlenecks while maintaining excellent chromatographic performance and mass spectrometry sensitivity.
Download this application note to explore:
- Ion-pairing-free methods
- Proven automated spectral deconvolution workflows
- Validated approaches for analyzing diverse oligonucleotides
Application Note
Biopharma
Authors
Lee Bertram and Brian Rivera
Agilent Technologies, Inc.
Abstract
This application note presents a practical approach for medium- to high-throughput
molecular confirmation of synthetic oligonucleotides using the Agilent InfinityLab
Pro iQ Plus LC/MS system. Traditional ion-pair reversed-phase LC/MS methods
often require dedicated instrumentation and toxic, costly reagents. In contrast, the
method described here uses an ammonium bicarbonate-based method, eliminating
the need for ion-pairing agents while maintaining sufficient chromatographic
retention and MS sensitivity. Agilent OpenLab CDS MS spectral deconvolution
simplifies data analysis workflows, enabling automated processing with minimal
optimization. This approach provides robust LC/MS performance for various
oligonucleotide types, including antisense oligonucleotides and siRNAs.
Analysis of Oligonucleotides Using
Ion-Pairing Alternatives on the
Agilent Pro iQ Plus
2
Introduction
Oligonucleotides are an emerging therapeutic modality
that target and modulate gene expression through the
silencing or degradation of mRNA. Once an mRNA target is
identified, the antisense sequence must be optimized with
strategically incorporated chemical modifications to improve
pharmacokinetics1
, affinity, and minimize off-targeting
or mismatching.2
Ion-pair reversed-phase LC/MS is commonly used to
confirm the molecular weights of target oligo sequences
to ensure proper synthesis. However, using alkylamines as
ion-pairing agents often requires dedicated instrumentation.
Furthermore, toxic and cost-prohibitive perfluorinated
alcohols such as HFIP (hexafluoroisopropanol) are required
for optimal chromatographic separation and MS sensitivity.
In this application, an alternative reversed-phase approach
was used for molecular confirmation of oligonucleotides. The
method used ammonium bicarbonate instead of ion pairing,
while still providing sufficient chromatographic retention
and MS sensitivity. Twenty replicate injections of three
different antisense oligonucleotides and five replicates of a
single siRNA were performed to ensure the applicability and
reproducibility of the method.
Experimental
Instrument configuration
This experiment was conducted using the following
instrument configuration:
– Agilent InfinityLab Pro iQ Plus LC/MS system (G6170A)
– Agilent Infinity II 1290 bio binary pump (G7120A)
– Agilent Infinity II 1290 bio multisampler (G7167B)
– Agilent Infinity II 1290 bio column compartment (G7116B)
– Agilent Infinity II 1260 diode array detector HS (G7117C)
Although this analysis used an Infinity II LC configuration,
comparable results can be achieved on the Infinity III LC
system with no changes to method parameters.
Sample preparation
All samples were resuspended in deionized (DI) water to
a concentration of 50 µM and stored at –80 °C. Samples
were transferred to polypropylene vials and stored in the
temperature-controlled autosampler for up to two days prior
to analysis. Oligonucleotide sample sequences are shown in
Table 1.
Oligonucleotide
Name Length Sequence
ASO-1 18
dU/MOErC//MOErA//MOErC/dUdUdU/MOErC//
MOErA/dU/MOErA//MOErA/dU/MOErG/CdU/
MOErG/G
ASO-2 20
dU/MOErC/dUdU/MOErG/TT/MOErA//MOErC//
MOErA//MOErT//MOErG//MOErA//MOErA//
MOErA/dU/MOErC//MOErC//MOErC/C
Fomivirsen 21 G*C*G*T*T*T*G*C*T*C*T*T*C*T*T*C*T*T*G*C*G
Givosiran
22 S mC*mA*mGmAmAmAfGmAfGmUfGmUfCmUfCmAm
UmCmUmUmA/L96/
23 AS mU*mG*mGfUmCfUmUfUfCmUfCfAmCfAmGfAmGfU
mAmGfA*fA*mU
Code Description
/MOErA/ Methoxyethoxy A
/MOErC/ Methoxyethoxy C
/MOErT/ Methoxyethoxy T
/MOErG/ Methoxyethoxy G
dU Deoxyuridine
fA 2-fluoroadenosine
fC 2-fluorocytidine
fG 2-fluoroguanadine
fU 2-fluorouridine
* Phosphorothioate bond
A 2'-deoxyribose adenine
C 2'-deoxyribose cytosine
G 2'-deoxyribose guanine
T 2'-deoxyribose thymine
mA 2'-O-methyl A
mC 2'-O-methyl C
mG 2'-O-methyl G
mU 2'-O-methyl U
rA Ribose adenine
rC Ribose cytosine
rG Ribose guanine
rU Ribose uracil
Table 1. Oligonucleotide sample sequences.
3
LC/MS analysis
Source parameters for the Pro iQ Plus system are provided
in Table 2, while high-performance liquid chromatography
(HPLC) parameters are provided in Table 3.
Mass Spectrometry Parameters
Parameter Value
MS Agilent Pro iQ Plus
Source Agilent Jet Stream Electrospray Ionization (AJS-ESI) Source
Drying Gas Flow 13.0 L/min
Gas Temp 300 °C
Nebulizer Pressure 35 psi
Capillary Voltage 3,000 V
Sheath Gas Temp 250 °C
Sheath Gas Flow 11 mL/min
Nozzle Voltage 1,500 V
Mode Positive
Scan m/z 700–2,800
Scan Time 1,250 ms
Fragmentor 180 V
Gain Factor 5
Table 2. Source parameters for the Agilent Pro iQ Plus system.
Parameter Value
Column Agilent AdvanceBio oligonucleotide column,
2.1 × 50 mm, 2.7 µm
Sampler Temperature 8 °C
UV Detection 260/4 nm (Ref 360/20 nm)
Peak width > 0.1 min (2.5 Hz)
Mobile Phase A 20 mM ammonium bicarbonate in DI water
Mobile Phase B Methanol
Flow Rate 0.7 mL/min
Injection Volume 2 µL
Multiwash 20:80 water:methanol; flush port; 5 seconds
90:10 water:methanol; flush port; 3 seconds
Column Temperature 75 °C
Post Time 1.0 min
Gradient Program
Time (min) %B
0 5
0.1 5
3.0 40
3.1 80
3.5 80
3.6 5
Table 3. HPLC parameters used.
Results and discussion
Established in 1997, an alkylamine ion pair with perfluorinated
alcohol as the acidic modifier is the preferred mobile phase
for LC/MS analysis of oligonucleotides.3
This is due to its
chromatographic performance and electrospray efficiency,
especially when compared to mobile phases using acetate
as the counter ion. Extensive work has further demonstrated
a wide experimental design space when using this powerful
ion-pairing system.4
However, there are consequences when using alkylamine
and HFIP for LC/MS methods. First, optimization of buffer
components is necessary since oligo modifications
and sequence affect electrospray desorption and
chromatography.5
Second, alkylamine may contaminate
ionization sources and LC systems, leading to background
peaks if polarity is switched back to positive mode.
Consequently, this requires extensive cleaning/passivation
of LC components and ion source surfaces. Even then, some
labs may dedicate systems to negative mode due to the
adsorption of alkylamines onto the LC/MS system. Finally,
alkylamine-containing mobile phases have a short shelf life
if not kept sealed under argon gas, and thus, mobile phases
must be made fresh, sometimes daily, to ensure consistent
method performance.
For labs performing molecular weight confirmation
workflows where many oligos are analyzed, optimizing
mobile phase conditions may not be feasible. Additionally,
LC/MS downtime due to instrument maintenance and
daily preparation of buffers may lead to sample backlogs,
which for many labs can be a considerable challenge. A
more practical approach would be to use a non-ion-pairing,
reversed-phase methodology.
Recent work provides a more practical and cost-effective
alternative to ion pairs, using an ammonium bicarbonate
(NH4
HCO3
) buffer and methanol as the strong solvent.6
This is
an advantageous method for molecular weight confirmation,
as ESI sensitivity and chromatographic performance are
sufficient, even with minimal optimization of the mobile phase
and gradient. Further, this referenced work postulates that
carbon dioxide outgassing facilitates droplet formation, while
ammonia evaporation contributes to proton adduction, thus
allowing for positive mode analysis. This, therefore, eliminates
the need to dedicate a system for negative mode analysis.
4
Because oligonucleotides are polar and negatively charged,
there may be concerns that these analytes would not be
retained or separate chromatographically without ion pairing.
However, ammonium bicarbonate-based methods retain
a wide variety of oligonucleotides and perform particularly
well with modified oligos, such as antisense oligos
(ASOs; Figure 1). The method demonstrated here provides a
proof-of-concept for medium- to high-throughput analysis.
Therefore, the gradient is 5% to 40% Mobile Phase B over
three minutes. This relatively "ballistic" gradient provides
excellent retention and peak shape, and the 0.7 mL/min
flow rate does not negatively impact the spectral quality
nor sensitivity.
Like most large molecules with multiple ionizable functional
groups, oligos appear as multiply charged ions when analyzed
by ESI-MS. Importantly, depending on the alkylamine ion
pair and source parameters, charge state distribution of
the oligo may vary.7
This can be problematic with heated
ionization sources, which tend to yield higher charge states.
Consequently, deconvolution of the spectrum may lead to
artifacts and potential misidentifications. On the other hand,
ammonium bicarbonate (ABC) tends to yield oligos with lower
charge states (Figure 2). The higher m/z values will have
less spectral overlap with matrix interferences, impurities,
or other components, thus improving confidence in the
deconvoluted data.
Figure 1. UV chromatograms for three antisense oligos (ASOs). Fomivirsen is a 21mer ASO with no 2′-modifications. Thus, it elutes earlier than the two shorter,
fully thioated, 2′-modified ASOs.
0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6
Retention time (min)
0
10
20
30
40
50
60
70
80
90
100
110
Relative response (%)
5
Agilent OpenLab CDS uses a unique spectral deconvolution
that is optimized for unit mass detectors. Maximum entropy
calculates the most probable mass from a spectrum by
increasing dominant features while minimizing noise. In
contrast, OpenLab CDS spectral deconvolution simply
identifies ions in the spectrum and matches them to target
masses based on user inputs. These values are then fit to
a linear regression or centroided to calculate a molecular
weight average. This simple approach allows for easy
molecular weight confirmation using generic processing
methods, which can be fully automated with minimal
optimization. Triplicate injections of each sample yield nearly
identical results, with relative standard deviations (%RSDs)
< 0.1%. The automated deconvolution settings used in the
sample sequence for acquisition are shown in Table 4, and a
summary of deconvolution results is shown in Table 5.
Parameter Value
Automatic Deconvolution Settings
Run Automatic Deconvolution Enabled
RT Window 1–2.7 min
TIC Peak Type All peaks
TIC Top (n) Peaks 3
Basic Settings
Use m/z Range Disabled
Low/High Molecular Weight 3,000–10,000
Maximum Charge 10
Minimum Peaks in Set 3
Advanced Settings
MW Agreement (0.01%) 10
Absolute Noise Threshold 100
Relative Abundance Threshold 10
MW Algorithm Curve fit
MW Algorithm Threshold (%) 40
Envelope Threshold (%) 50
Table 4. MS spectral deconvolution settings.
Figure 2. Full-scan spectra for samples. Charge states of 4 to 6 are predominant for each ASO.
800 1,000 1,200 1,400 1,600 1,800 2,000 2,200 2,400 2,600 2,800
m/z
0
1
2
3
Abundance
703.1 834.9 914.7 1,026.7
1,045.1
1,113.8
1,197.6
1,219.1
1,286.3 1,440.1
1,462.7
1,598.6 1,733.4
1,828.2
1,954.6 2,088.5 2,179.2 2,337.3 2,437.3 2,583.2 2,682.4
Max: 383,860
800 1,000 1,200 1,400 1,600 1,800 2,000 2,200 2,400 2,600 2,800
m/z
0
1
Abundance
709.5 822.4 907.5 1,033.7
1,059.0
1,128.9
1,240.4
1,270.3
1,287.7 1,416.01,501.2
1,587.7
1,589.3 1,716.5 1,874.4 1,993.4 2,117.3 2,250.8 2,343.1 2,457.0
Max: 222,126
800 1,000 1,200 1,400 1,600 1,800 2,000 2,200 2,400 2600 2,800
m/z
0
2
4
Abundance
721.2 835.0 956.31,041.3
1,089.7
1,114.5
1,115.3
1,124.31,212.6
1,306.2
1,336.9
1,338.0
1404.4 1,489.2 1,584.5
1,670.6
1,671.6
1,672.8
1,691.4 1,781.6 1,873.9 1,969.6 2,063.8 2,226.0
×10 Max: 63,956 4
×105
×105
6
Figure 3 shows the deconvoluted spectrum of ASO-1,
an 18mer phosphorothioate with 2′-MOE modifications.
Component A is a full-length product, showing good
agreement with the calculated mass. Component B is likely
depurination (loss of guanine). Component C is likely a
sodium adduct. As shown, the InfinityLab Pro iQ Plus, which
combines unit mass accuracy with simple deconvolution, can
monitor oligonucleotides and identify unexpected impurities
or components. Further characterization can be performed by
high-resolution MS.
Oligonucleotide
Name
Calculated Mass
(Da)
Average Measured
Mass (Da) Delta Mass (Da)
ASO-1 6,348.3 6,347.9 –0.4
ASO-2 7,309.2 7,308.8 –0.4
Fomivirsen 6,682.4 6,681.8 –0.6
Table 5. Deconvoluted results of twenty replicates of each ASO.
A duplex, tri-antennary GalNAc-conjugated siRNA (Givosiran)
was also analyzed to demonstrate broad applicability of the
method. Slight modifications to the LC/MS parameters were
required with the duplex conjugate. Specifically, the gradient
program was adjusted slightly (5% to 45% in 3 minutes)
to ensure baseline separation of the sense and antisense
strands. Additionally, the fragmentor voltage was decreased
to 120 V to minimize in-source fragmentation of the fragile
oligosaccharide conjugate. Figure 4 shows an overlay of UV
chromatograms for five injections of Givosiran.
Although no ion pair is used, the ABC mobile phase still
provides baseline separation of sense and antisense strands
(Figure 5). This separation is required because any spectral
overlap may result in misidentifications in deconvolution.
Spectra are shown in Figure 5, with similar charge state
distributions to the previously mentioned ASOs. Analysis of
the deconvoluted spectrum for each peak show the antisense
strand eluting earlier than the sense strand. Interestingly,
the sense strand may indicate GalNAc losses (Figure 6). A
summary of Givosiran results is shown in Table 6.
Oligonucleotide Name
Calculated Mass
(Da)
Average Measured
Mass (Da) Delta Mass (Da)
Givosiran, Antisense 7,563.8 7,563.1 0.7
Givosiran, Sense 8,736.5 8,735.6 0.9
Table 6. Deconvoluted results for each injection of Givosiran.
Figure 3. Deconvoluted spectrum for ASO-1.
5,800 5,900 6,000 6,100 6,200 6,300 6,400 6,500 6,600 6,700 6,800 6,900 7,000
Mass (Daltons)
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
Abundance
A:6,347.88
B:6,196.24
C:6,369.33
×105
7
Figure 4. Overlay of UV chromatograms for five injections of Givosiran.
0.55 0.60 0.65 0.70 0.75 0.80 0.85 0.90 0.95 1.00 1.05 1.10 1.15 1.20 1.25 1.30 1.35 1.40 1.45 1.50 1.55 1.60 1.65 1.70
Retention time (min)
0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
Response (mAU)
Integration Integration
×101
Figure 5. Spectra for antisense (A) and sense (B) RNA strands. Charge state distribution is similar to that of antisense oligos analyzed using a similar method. The
5– charge state is the most abundant for antisense, and the 6– charge state is the most abundant for the GalNAc-conjugated sense strand.
800 1,000 1,200 1,400 1,600 1,800 2,000 2,200 2,400 2,600 2,800
m/z
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
2.2
Abundance
761.7 843.5 946.3
953.41,039.9
1,081.1
1,087.5
1,151.3
1,254.1
1,261.7
1,265.2
1,283.0 1,422.5
1,508.7
1,513.7
1,517.8
1,519.6
1,521.9
1,613.6 1,718.91,802.11,889.8
1,891.4
1,968.9 2,122.9 2,212.8 2,313.12,397.0 2,536.02,621.62,703.32,780.3
Max: 213,395
800 1,000 1,200 1,400 1,600 1,800 2,000 2,200 2,400 2,600 2,800
m/z
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
Abundance
711.9
776.7
797.5
854.2
874.5
877.0
949.1
951.4
952.7
971.6
974.0
1,067.4
1,081.1
1,092.9
1,095.6
1,098.2
1,176.0
1,219.9
1,222.9
1,248.9
1,252.2
1,255.8
1,352.6
1,406.4
1,422.9
1,430.1 1,626.41,707.7
1,810.6 2,043.02,130.1 2,248.6 2,343.8 2,440.8 2,605.8 2,776.4
Max: 1,101,977
×105
×106
A
B
8
Figure 6. Deconvoluted spectrum for the Givosiran sense strand. The approximately 200 Da shift observed could be loss of GalNAc.
7,800 8,000 8,200 8,400 8,600 8,800 9,000 9,200 9,400
Mass (Daltons)
0
0.5
1.0
1.5
2.0
2.5
Abundance
A:8,735.42
B:8,532.26
C:8,758.10
E:8,328.63 D:8,554.26
×106
Conclusion
The method described in this application note provides a
practical approach for medium- to high-throughput molecular
confirmation of synthetic oligonucleotides. The Agilent Pro
iQ Plus demonstrates excellent resolution and sensitivity for
a unit mass detector, even at high m/z ranges that exceed
the capabilities of many single quadrupole detectors. The
method uses a novel ammonium bicarbonate mobile phase,
providing sufficient LC/MS performance for labs analyzing
many types of oligos, including antisense oligonucleotides
and siRNAs. LC and MS results described in this application
note are reproducible. Data analysis workflows are simplified
using Agilent OpenLab CDS MS spectral deconvolution, which
is automated by unattended processing methods that require
minimal optimization.
www.agilent.com
DE-006128
This information is subject to change without notice.
© Agilent Technologies, Inc. 2025
Printed in the USA, May 6, 2025
5994-8337EN
References
1. Herkt, M.; Thum, T. Pharmacokinetics and Proceedings
in Clinical Application of Nucleic Acid Therapeutics.
Mol. Ther. 2021, 29(2), 521–539. DOI: 10.1016/j.
ymthe.2020.11.008.
2. Watts, J. K.; Corey, D. R. Silencing Disease Genes in
the Laboratory and the Clinic. J. Pathol. 2012, 226(2),
365–379. DOI: 10.1002/path.2993.
3. Apffel, A.; Chakel, J. A.; Fischer, S.; Lichtenwalter, K.;
Hancock, W. S. Analysis of Oligonucleotides by HPLC−
Electrospray Ionization Mass Spectrometry. Anal. Chem.
1997, 69(7), 1320–1325. DOI: 10.1021/ac960916h.
4. Guimaraes, G. J.; Bartlett, M. G. The Critical Role of Mobile
Phase pH in the Performance of Oligonucleotide Ion-Pair
Liquid Chromatography-Mass Spectrometry Methods.
Future Sci. OA 2021, 7(10), FSO753. DOI: 10.2144/fsoa2021-0084.
5. Basiri, B.; Murph, M. M.; Bartlett, M. G. Assessing the
Interplay Between the Physicochemical Parameters of
Ion-Pairing Reagents and the Analyte Sequence on the
Electrospray Desorption Process for Oligonucleotides.
J. Am. Soc. Mass Spectrom. 2017, 28(8), 1647–1656. DOI:
10.1007/s13361-017-1671-6.
6. Hayashi, Y.; Sun, Y. Overcoming Challenges in
Oligonucleotide Therapeutics Analysis: A Novel Nonion
Pair Approach. J. Am. Soc. Mass Spectrom. 2024, 35(9),
2034–2037. DOI: 10.1021/jasms.4c00270.
7. Chen, B.; Mason, S. F.; Bartlett, M. G. The Effect
of Organic Modifiers on Electrospray Ionization
Charge-State Distribution and Desorption Efficiency for
Oligonucleotides. J. Am. Soc. Mass Spectrom. 2013,
24(2), 257–264. DOI: 10.1007/s13361-012-0509-5.
Brought to you by
Download the Application Note for FREE Now!
Information you provide will be shared with the sponsors for this content. Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version