Optimize Oligonucleotide Analysis With Advanced Chromatography Methods
App Note / Case Study
Published: July 10, 2025
Credit: iStock
Oligonucleotide therapeutics are revolutionizing modern medicine, but their complex analytical characterization remains a significant challenge for biopharma laboratories.
Traditional ion-pairing reversed-phase liquid chromatography methods require dedicated instruments and present compatibility issues with mass spectrometry (MS) detection.
This application note reveals how hydrophilic interaction chromatography (HILIC) provides a versatile, MS-compatible alternative that delivers superior separation performance without the limitations of conventional approaches.
Download this application note to learn:
- How to select and optimize HILIC stationary phases for different oligonucleotide types and modifications
- Critical mobile phase pH considerations that impact charge state distribution and ionization efficiency
- Proven methods for achieving baseline separation of challenging oligonucleotide pairs
Application Note
Biopharma/Pharma
Authors
Jordy Hsiao,
Lee Bertram, Alex Apffel,
Andrea Angelo P. Tripodi,
Andrew Coffey, Ta-Chen Wei,
and Connor Flannery
Agilent Technologies, Inc.
Abstract
Hydrophilic interaction chromatography (HILIC) represents a versatile and effective
alternative to ion-pairing reversed-phase liquid chromatography (IP-RPLC) for
analyzing oligonucleotides. In this study, five different HILIC stationary phases
were comprehensively evaluated and optimized to determine their suitability and
complementarity for oligonucleotide analysis using liquid chromatography/mass
spectrometry (LC/MS). Key factors for optimizing retention and detection include
column selectivity and the pH of the mobile phase. Further studies were then
carried out to characterize heavily modified oligonucleotides with the optimized
HILIC-LC/MS method.
Evaluating HILIC Stationary Phases
for Oligonucleotide Separation
by LC/MS
2
Introduction
IP-RPLC coupled with MS represents
the most common analytical method
for oligonucleotide analysis.1-3 However,
alternative separation methods are
needed because alkylamine ion-pair
reagents require dedicated instruments.
While ion-exchange chromatography
(IEX) represents a viable alternative
technique due to excellent selectivity for
oligonucleotides based on their length,
it is not preferred due to mobile phase
incompatibility with MS detection. HILIC
is a valuable alternative to both IP-RPLC
and IEX as HILIC mobile phases are
compatible with MS and offer flexibility
in instrument use. This work highlights
the utility of HILIC for oligonucleotide
analysis and the critical parameters
that need to be considered to optimize
LC/MS performance.
Experimental
Reagents and supplies
– DNA ladder standard
(part number 5190-9029)
– RNA resolution standard
(part number 5190-9028)
– Polypropylene vials
(part number 5188-2788)
– Blue snap caps
(part number 5182-3458)
Sample preparation
Oligonucleotide standards covering
DNA and RNA sequences ranging from
14- to 40-mer in length were selected for
HILIC stationary phase screening and
HILIC liquid chromatography/quadrupole
time of flight (LC/Q-TOF) MS method
development. A pharmaceutically
relevant 18-mer antisense
oligonucleotide (ASO) was then used
to validate the optimized HILIC-MS
methods. All samples were resuspended
in nuclease-free water and stored
at –20 °C. The samples were then
transferred to polypropylene vials
and kept in a temperature‑controlled
autosampler for up to 2 days
before analysis. The sequences and
modifications for the 14-, 17-, 20-, and
21-mer RNA and the 18-mer ASO can
be found in a previously published
application note.4
HILIC columns
Five HILIC columns from Agilent were
evaluated in this study: InfinityLab
Poroshell 120 HILIC (P120 HILIC),
InfinityLab Poroshell 120 HILIC-OH5,
InfinityLab Poroshell 120 HILIC-Z,
AdvanceBio Amide HILIC, and
AdvanceBio Glycan Mapping. The Glycan
Mapping column features a neutral
amide stationary phase, while the Amide
HILIC column has a mixed-mode HILIC
stationary phase with both amide and
ion-exchange functionalities. All columns
contain 2.7 μm superficially porous
particles, except for the Amide HILIC
column, which has 1.8 μm fully porous
particles. A summary of the columns
used is provided in Table 1.
Mobile phase preparation
LC/MS-grade acetonitrile and Milli-Q
purified water were used for the LC/MS
analysis. Detailed steps for solvent buffer
preparation are shown in Table 2.
Table 1. List of HILIC stationary phases evaluated for oligonucleotide separation.
No. Column
Pore Size
(Å)
Dimensions
(mm)
Particle Size
(μm) Surface Chemistry pH Range
Pressure
Rating (bar)
Maximum
Temperature (°C) Part Number
1 HILIC 120 2.1 × 150 2.7 Bare silica 0 to 8 600 – 693775-901
2 HILIC-OH5 120 2.1 × 150 2.7 Poly-hydroxy fructan 1 to 7 400 45 683775-601
3 HILIC-Z 100 2.1 × 150 2.7 Zwitterionic 2 to 12 600 80 683775-924
4 Glycan Map 120 2.1 × 150 2.7 Neutral amide 2 to 7 600 40 683775-913
5 Amide HILIC 300 2.1 × 150 1.8 Mixed mode amide 2 to 7 1,200 80 859750-91
Table 2. Mobile phase preparation for LC/MS analysis.
Method 10x Stock Buffer Mobile Phase A Mobile Phase B
High pH Method 100 mM ammonium acetate, pH 9.0,
adjusted with ammonium hydroxide
100 mL of stock buffer was
mixed with 900 mL of water
100 mL of stock buffer was
mixed with 900 mL of ACN Neutral pH Method 100 mM ammonium acetate, pH 6.8
Low pH Method 100 mM ammonium acetate, pH 4.4,
adjusted with acetic acid
3
Instrumentation
LC/MS analysis was performed using an
Agilent 1290 Infinity II LC system coupled
to an Agilent 6545XT AdvanceBio
LC/Q-TOF with an Agilent Jet Stream
Electrospray source. The LC consisted of:
– Agilent 1290 Infinity II high-speed
pump (G7120A)
– Agilent 1290 Infinity II multisampler
with thermostat (G7167B)
– Agilent 1290 Infinity II multicolumn
thermostat (MCT) (G7116B)
– Agilent 1290 Infinity II diode array
detector (DAD) (G7117A)
Dynamic mass axis correction was
achieved by continuous infusion of a
reference mass solution. Tables 3 and 4
summarize the optimized LC and the MS
conditions. Data acquisition and analysis
was done using the Agilent MassHunter
software suite.
Table 3. LC parameters for 18-mer ASO analysis.
Agilent 1290 Infinity II LC System
Column
Agilent lnfinityLab Poroshell 120 HILIC-Z,
2.1 mm × 150 mm, 2.7 μm
(part number 683775-924)
Agilent AdvanceBio Glycan Mapping 120Å,
2.1 mm × 150 mm, 2.7 μm
(part number 683775-913)
Solvent
A) 10 mM ammonium acetate in water,
pH 9.0
B) 10 mM ammonium acetate in
water/ACN 10:90 (v:v), pH 9.0
A) 10 mM ammonium acetate in water,
pH 6.8
B) 10 mM ammonium acetate in
water/ACN 10:90 (v:v), pH 6.8
Gradient
Time (min) %B
0.00 80
2.00 80
9.00 60
11.0 60
12.0 80
17.0 80
Column Temperature 30 °C
Flow Rate 0.25 mL/min
Autosampler Temperature 8 °C
Table 4. MS parameters for 18-mer ASO analysis.
Agilent 6S45XT AdvanceBio LC/Q-TOF
Ionization Mode Dual AJS ESI
Ionization Polarity Negative
Gas Temperature 275 °C
Drying Gas 12 L/ min
Nebulizer Pressure 35 psi
Sheath Gas temperature 350 °C
Sheath Gas Flow 12 L/min
Capillary Voltage 3,500 V
Nozzle Voltage 2,000 V
Fragmentor 175 V
Skimmer 65 V
Octopole 1 RF Voltage 750 V
Acquisition Range m/z 300 to 3,200
MS Acquisition rate 1 spectrum/s
Reference Mass m/z 980.01637500
4
Results and discussion
To assess the utility of different HILIC
stationary phases for oligonucleotide
separation, the columns were evaluated
with a mobile phase containing 10 mM
ammonium acetate buffers at pH 4.4,
6.8, and 9.0. The varying mobile
phase compositions could change the
stationary phase selectivity, thereby
altering oligonucleotide separation.
Moreover, mobile phase buffer pH may
influence the electrospray ionization of
oligonucleotides.5
Understanding and
balancing the impact of buffer choice
on retention, selectivity, and ionization
efficiency is critical when developing
optimal LC/MS methods.
To ensure a fair comparison across
all columns, the same gradient slope
(30% B change in 10 minutes, 3% B/min)
was used. In general, the columns
performed better at pH 6.8 than the
acidic pH 4.4 condition (Figure 1). This
is understandable given the benefits of
elevated pH for oligonucleotide analysis
because it promotes conditions that
result in better retention, resolution,
selectivity, and ionization for IP‑RP.6
Moreover, the Glycan Mapping column
yielded the best selectivity in resolving
the varied oligonucleotides of different
sequence lengths, followed by the
HILIC-Z, then the HILIC-OH5 columns.
The P120 HILIC bare silica column
showed poor selectivity for the DNA
oligonucleotides, while the Amide
HILIC column results were unexpected,
as changes to the mobile phase
composition did not yield any detectable
oligonucleotide peak. This may be due
to the ion-exchange properties of the
mix-mode column that are reflective
of the structural complexity of the
oligonucleotides as well as the positional
location of the ion exchange functionality
on the different bonded HILIC phases.
Further mobile phase and gradient
optimizations such as increase in salt
concentrations are required to elute the
oligonucleotides off the column.
Figure 1. HILIC-MS analysis of 15-, 20-, 25-, 30-, 35-, and 40-mer DNA samples using HILIC stationary phases with varying chemical properties with 10 mM
ammonium acetate buffer at (A) pH 4.4 and (B) pH 6.8. The gradient used for each column is listed in the top right corner of each chromatogram. The gradient
was modified to improve peak resolution and adjust for retention time shifts when switching the pH of the mobile phase.
5
Given the improved performance of the
oligonucleotides going from acidic to
neutral pH, the impact of elevated pH
was further explored with the HILIC-Z
column because of its stability at higher
pH.7
The oligonucleotide sample was
analyzed with the HILIC-Z column
at pH 6.8 and pH 9.0 (Figure 2). At
pH 6.8, the HILIC-Z column yielded
slightly broader peaks than pH 9.0 with
lower resolution between the 35- and
40‑mer oligonucleotides. At pH 9.0, the
oligonucleotide peaks were noticeably
sharper. Baseline separation could be
achieved by flattening the gradient from
3 to 2% B/min, with the potential to use
an even shallower gradient to further
resolve the oligonucleotides.
Based on the results shown in
Figure 2, it was decided to continue
the downstream analysis with the
Glycan Mapping column at pH 6.8 and
the HILIC-Z column at pH 9.0. RNA
samples composed of 14-, 17-, 20- and
21-mer lengths were analyzed on the
two columns (Figure 3). Importantly,
the 20- and 21-mer RNA samples were
resolved with both columns, where
separation of the n–1 oligonucleotide
pair represents a critical attribute that
is closely monitored for the incomplete
synthesis of an oligonucleotide.8,9
Figure 2. HILIC separation of 15-, 20-, 25-, 30-, 35-, and 40-mer DNA using an Agilent HILIC-Z column at
pH 6.8 and pH 9.
Figure 3. HILIC separation of 14-, 17-, 20-, 21-mer RNA using an Agilent Glycan Mapping column at pH 6.8
(A) and an Agilent HILIC-Z column at pH 9 (B).
A
B
6
A closer examination of the mass
spectra revealed a wider charge state
distribution for the oligonucleotides
analyzed with mobile phase buffer at
pH 6.8 than that at pH 9.0. Specifically,
a range of 4- to 8- charge states were
observed for the 21-mer RNA at pH 6.8
(Figure 4A). In contrast, a range of
4- to 6- charge states were observed
for the RNA oligonucleotides at pH 9.0
(Figure 4B). As the length of the
oligonucleotide sequence increases, the
ability to detect higher charge states
becomes necessary for detection within
the mass range limitations of the mass
spectrometer. Furthermore, a broader
range of detectable charge states
lends more confidence in the target
analyte identification.
Figure 4. Full scan MS1 mass spectra for 14-, 17-, 20-, 21-mer RNA.
7
In addition to optimizing the
chromatographic conditions for the
columns, it is equally important to
determine the method compatibility to
yield an optimal MS signal response. This
optimization allows users to confidently
identify the peaks detected by LC/MS.
This is demonstrated in Figure 5, where
the spectral deconvolution of the mass
spectra confirmes the masses for the
14-, 17-, 20-, and 21-mer RNA.
Figure 5. Spectral deconvolution of 14-, 17-, 20-, 21-mer RNA, showing the applicability of the LC/MS method for intact mass determination.
8
To evaluate the optimized HILIC-MS
methods with a heavily modified
oligonucleotide, an ASO with multiple
2-methoxyethoxy (2MOE) modifications
were analyzed using the Glycan Mapping
and HILIC-Z column (Figure 6). The
results showed that both columns
yielded a sharp peak for the ASO, with the
Glycan Mapping column having better
sensitivity relative to the HILIC-Z column
(Figure 6A). As expected, the Glycan
Mapping also showed a broader charge
state distribution with the 8- charge state
being the highest charge state detected.
In contrast, the HILIC-Z column highest
charge state detected was 6- for the ASO
(Figure 6B). Lastly, the ASO identity was
confirmed through the deconvoluted
mass spectra from both experiments
(Figure 6C).
The stability of the Glycan Mapping
column was verified by running and
storing it with 10 mM ammonium
acetate at pH 6.8 and 30 °C over 17 days
(Figure 7). The column underwent
repeated RNA sample injections for
3 days, paused for 2 days, run for
5 days, paused for 3 days, then run
for an additional 5 days to complete
1,000 injections. This stop/start
pattern simulates a more realistic
usage scenario for oligonucleotide
analysis. The chromatographic
results demonstrated that the column
maintained optimal separation
between the n–1 oligonucleotide pair
(i.e., change in retention time (ΔRT)
between 20- and 21-mer) throughout the
1,000 injections. Additionally, the method
achieved excellent retention stability,
with a %RSD of less than 3% for all
monitored oligonucleotides.
Figure 6. Comparison of the Glycan Mapping and HILIC-Z columns for characterization of an 18-mer ASO.
Figure 7. Reproducibility study with the Glycan Mapping column using an Agilent 1290 Infinity diode array detector over 1,000 injections.
www.agilent.com
DE-004781
This information is subject to change without notice.
© Agilent Technologies, Inc. 2025
Printed in the USA, March 17, 2025
5994-8180EN
Conclusion
This study systematically evaluated
five different HILIC stationary phases
with varying mobile phase pH levels to
determine the optimal conditions for
chromatographic peak shape, resolution,
and MS signal for oligonucleotides. It
was found that different mobile phase
pH conditions can impact the charge
state distribution of the analyte, which is
important to consider when working with
longer oligonucleotides. The developed
methods ultimately facilitated the LC/MS
analysis of a heavily modified ASO, with
the potential for sequence confirmation
analysis through tandem MS. This work
confirms that HILIC chromatography
can be an attractive alternative to IP‑RP
for the analytical characterization
of oligonucleotides. These methods
could be applied to other emerging
oligonucleotide modalities, including
siRNA, aptamers, single-stranded guide
RNA (sgRNA), and mRNA sequencing in
the future.
References
1. El Zahar, N. M.; Magdy, N.;
El‑Kosasy, A. M.; Bartlett, M. G.
Chromatographic Approaches for the
Characterization and Quality Control
of Therapeutic Oligonucleotide
Impurities. Biomed. Chromatogr.
2018, 32(1). doi: 10.1002/bmc.4088.
2. Goyon, A.; Yehl, P.; Zhang, K.
Characterization of Therapeutic
Oligonucleotides by Liquid
Chromatography. J. Pharm.
Biomed. Anal. 2020, 182, 113105.
doi: 10.1016/j.jpba.2020.113105.
3. Roussis, S. G.; Koch, C.; Capaldi, D.;
Rentel, C. Rapid Oligonucleotide Drug
Impurity Determination by Direct
Spectral Comparison of Ion-Pair
Reversed-Phase High-Performance
Liquid Chromatography Electrospray
Ionization Mass Spectrometry Data.
Rapid Commun. Mass Spectrom.
2018, 32(14), 1099–1106. doi:
10.1002/rcm.8125.
4. Bertram, L; Hsiao, J. Analysis of
Oligonucleotides Using an IonPairing-Free Reversed-Phase
Method with TOF LC/MS, Agilent
Technologies application note,
publication number 5994‑8013EN,
2024.
5. Apffel, A.; Chakel, J. A.; Fischer, S.;
Lichtenwalter, K.; Hancock. W. S.
New Procedure for the Use
of High-Performance Liquid
Chromatography–Electrospray
Ionization Mass Spectrometry for
the Analysis of Nucleotides and
Oligonucleotides. J. Chromatogr A
1997, 777(1), 3–21. doi: 10.1016/
S0021-9673(97)00256-2
6. Guimaraes, G. J. and Bartlett, M. G.
The Critical Role of Mobile
Phase pH in the Performance
of Oligonucleotide Ion-Pair
Liquid Chromatography–Mass
Spectrometry Methods. Future
Sci. OA 2021, 7(10), FSO753.
doi: 10.2144/fsoa-2021-0084.
7. Hsiao, J. J.; Kennedy, A. P.;
Van de Bittner, G. C.; Wei, T. The
Use of HILIC Zwitterionic Phase
Superficially Porous Particles for
Metabolomics Analysis. LCGC
Supplements Special Issues 2018,
36(6), 30–35.
8. Temsamani, J.; Kubert, M.;
Agrawal, S. Sequence Identity of
the n -1 Product of a Synthetic
Oligonucleotide. Nucleic Acids Res.
1995, 23(11), 1841–1844. doi:
10.1093/nar/23.11.1841.
9. Fearon, K. L.; Stults, J. T.;
Bergot, B. J.; Christensen, L. M.;
Raible, A. M. Investigation of the
'n–1' Impurity in Phosphorothioate
Oligodeoxynucleotides Synthesized
by the Solid-Phase β-Cyanoethyl
Phosphoramidite Method Using
Stepwise Sulfurization. Nucleic
Acids Res. 1995, 23(14), 2754–2761
(1995). doi: 10.1093/nar/23.14.2754.
Brought to you by
Download the Application Note for FREE Now!
Information you provide will be shared with the sponsors for this content. Technology Networks or its sponsors may contact you to offer you content or products based on your interest in this topic. You may opt-out at any time.
Experiencing issues viewing the form? Click here to access an alternate version